Recall Register

Food, drugs, devices › Tornier, Inc

Device recall: Aequalis Ascend Flex Humeral Stem: Part Number UDI DWF601A 3700386944475 DWF601B 3700386944482 DWF601C 3700386944499 DWF

FDA enforcement event 84033 · Class II · initiated 18 Oct 2019 · status Terminated · terminated 9 Apr 2021

DeviceVoluntary: Firm initiated1 product recordBloomington, MN

FDA enforcement event 84033 is a device recall by Tornier, Inc of Bloomington, MN, initiated 18 Oct 2019 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 968 device enforcement events initiated in 2019 carry the same class.

Status: Terminated, terminated 9 Apr 2021. Initiation: voluntary: firm initiated. Tornier, Inc has 17 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Tornier, Inc, Bloomington, MN.

Reason for recall

"During shipping, it is possible for the Aequalis Ascend Flex humeral stem to become unseated or otherwise move within its internal packaging restraints. Movement within the packaging may cause the product to rub against the PETG blister shipping container causing the transfer of PETG material from the blister container to the stem."

Distribution

"US Nationwide distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0405-2020Aequalis Ascend Flex Humeral Stem: Part Number UDI DWF601A 3700386944475 DWF601B 3700386944482 DWF601C 3700386944499 DWF602A 3700386944505 DWF602B 3700386944512 DWF602C 3700386944529 DWF603A 3700386944536 DWF603B 3700386944543 DWF603C 3700386944550 DWF604A 3700386944567 DWF604B 3700386944574 DWF605A 3700386944598…87869 unitsClass IITerminated20 Nov 2019

Context

88%of 2019 device events are Class II968 events that year
1product record in this eventfirm median: 1
17events on this register for the firm

Other enforcement events for Tornier, Inc

InitiatedProduct (first record)TypeClassRecords
5 Mar 2026Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max TuberoDeviceClass II1
19 Dec 2025TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318DeviceClass II1
3 Mar 2025stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103; Prosthesis, Elbow, SemDeviceClass II1
29 Aug 2024stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the sDeviceClass II1
29 Aug 2024Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplastyDeviceClass II1
3 Jan 2024Stryker Tornier Perform Reversed Glenoid Press-Fit Short Post Ti6AI4V, REF DWJ001, L: 7mm, Cementless Sterile.DeviceClass II1
15 Sep 2023stryker Tornier Perform Reversed Augmented Glenoid Lateralized Baseplate, 25mm, Offset +3mm, REF DWJ502DeviceClass II1
12 Jun 2023AEQUALIS FLEX REVIVE Assembly Screw 0mm, Model Number ARS655101DeviceClass II2
3 May 2022Ratcheting Screwdriver Handle, provided in the Aequalis Perform Reversed kitDeviceClass II1
28 May 2021Aequalis Ascend Flex Shoulder System Reversed Platform (Reversed Tray component) Rx Only Sterile Manufacturer Tornier SADeviceClass II1

All 17 events for Tornier, Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "During shipping, it is possible for the Aequalis Ascend Flex humeral stem to become unseated or otherwise move within its internal packaging restraints. Movement within the packaging may cause the product to rub against the PETG blister shipping container causing the transfer of PETG material from the blister container to the stem."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 9 Apr 2021. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.