Recall Register

Food, drugs, devices › Isopure Corp

Device recall: Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #11019

FDA enforcement event 83990 · Class II · initiated 19 Apr 2019 · status Terminated · terminated 29 May 2020

DeviceVoluntary: Firm initiated3 product recordsLouisville, KY

Isopure Corp (Louisville, KY) initiated this device recall on 19 Apr 2019; FDA lists 3 product records under event 83990, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 968 device enforcement events initiated in 2019 carry the same class.

Status: Terminated, terminated 29 May 2020. Initiation: voluntary: firm initiated. Isopure Corp has 3 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Isopure Corp, Louisville, KY.

Reason for recall

"There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron."

Distribution

"The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0821-2020Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-315Class IITerminated29 Jan 2020
Z-0822-2020Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-31Class IITerminated29 Jan 2020
Z-0823-2020Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)8Class IITerminated29 Jan 2020

Context

88%of 2019 device events are Class II968 events that year
3product records in this eventfirm median: 1
3events on this register for the firm

Other enforcement events for Isopure Corp

InitiatedProduct (first record)TypeClassRecords
26 Jul 2022Acidified Concentrate Distribution and Storage-The Isopure ACDS (Acidified Concentrate Distribution and Storage) with reDeviceClass II1
28 Apr 2022Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO)DeviceClass II1

All 3 events for Isopure Corp

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 29 May 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.