Recall Register

Food, drugs, devices › Beckman Coulter Inc.

Device recall: Beckman Coulter Lipase, REF OSR6130 and OSR6230.

FDA enforcement event 83583 · Class II · initiated 24 Aug 2018 · status Terminated · terminated 2 Aug 2024

DeviceVoluntary: Firm initiated2 product recordsBrea, CA

Beckman Coulter Inc. (Brea, CA) initiated this device recall on 24 Aug 2018; FDA lists 2 product records under event 83583, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 92% of the 1,074 device enforcement events initiated in 2018 carry the same class.

Status: Terminated, terminated 2 Aug 2024. Initiation: voluntary: firm initiated. Beckman Coulter Inc. has 216 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Beckman Coulter Inc., Brea, CA.

Reason for recall

"NAPQI, a metabolite of Acetominophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients."

Distribution

"Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0210-2020Beckman Coulter Lipase, REF OSR6130 and OSR6230.20,926 unitsClass IITerminated30 Oct 2019
Z-0211-2020Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698.19,834 uitsClass IITerminated30 Oct 2019

Context

92%of 2018 device events are Class II1,074 events that year
2product records in this eventfirm median: 1
216events on this register for the firm

Other enforcement events for Beckman Coulter Inc.

InitiatedProduct (first record)TypeClassRecords
15 Jul 2026Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC DeviceClass II2
26 Jun 2026REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-28DeviceClass II2
16 Mar 2026Access Total T4 Calibrator, Catalog No. 33805DeviceClass II1
23 Dec 2025CellMek SPS Sample Preparation System, REF: C44603, with softwareDeviceClass II1
26 Nov 2025Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin SubunitDeviceClass II2
10 Nov 2025COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunctioDeviceClass II1
7 Nov 2025Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibratioDeviceClass II1
1 Oct 2025Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry AnalyzerDeviceClass II2
22 Sep 2025DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use asDeviceClass II1
19 Sep 2025DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i DeviceClass II2

All 216 events for Beckman Coulter Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "NAPQI, a metabolite of Acetominophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 2 Aug 2024. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.