Recall Register

Food, drugs, devices › GE Healthcare, LLC

Device recall: Aestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077

FDA enforcement event 83425 · Class II · initiated 19 Jul 2019 · status Terminated · terminated 13 Apr 2026

DeviceVoluntary: Firm initiated4 product recordsWaukesha, WI

FDA enforcement event 83425 is a device recall by GE Healthcare, LLC of Waukesha, WI, initiated 19 Jul 2019 and classified Class II, with 4 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 968 device enforcement events initiated in 2019 carry the same class.

Status: Terminated, terminated 13 Apr 2026. Initiation: voluntary: firm initiated. GE Healthcare, LLC has 225 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

GE Healthcare, LLC, Waukesha, WI.

Reason for recall

"Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters."

Distribution

"Worldwide"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0114-2020Aestiva MRI, Model Numbers: a) 1006-9310-000 b) 1006-9110-000 c) 1006-9023-000 d) 1006-9028-000 e) 1006-9310-000-305077 f) 1006-9310-000-015243 g) 1006-9310-000-017602 h) 1006-9310-000-103785 I) 1006-9310-000-025109 j) 1006-9310-000-009650 k) 1006-9310-000-015224 l) 1006-9310-000-031881 m) 1006-9310-000-031854 n)…502 devicesClass IITerminated25 Dec 2019
Z-0115-2020Aestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n) 1006-9120-000 o) 1006-9027-000 p) 1006-9003-000 q) 1006-9025-0007168 devicesClass IITerminated25 Dec 2019
Z-0116-2020Aespire 7100, Model Numbers: 1. 1009-9000-000-009469 2. 1009-9000-000 3. 1009-9000-000-013913 4. 1009-9000-000-001366 5. 1009-9000-000-009857 6. 1009-9003-000 7. 1009-9000-000-008120 8. 1009-9000-000-009600 9. 1009-9000-000-014290 10. 1009-9000-000-010585 11. 1009-9000-000-004993 12. 1009-9011-000-004795 13.…19351 devicesClass IITerminated25 Dec 2019
Z-0117-2020Aestiva 7100, Model Numbers: 1. 1006-9320-000 2. 1006-9321-000 3. 1006-9320-000-001101 4. 1006-9021-000 5. 1006-9012-000 6. 1006-9111-000 7. 1006-9321-000-007639 8. 1006-9321-000-000023 9. 1006-9321-000-015180 10. 1006-9320-000-010369 11. 1006-9321-000-019486 12. 1006-9321-000-007601 13. 1006-9321-000-009266 14.…473 devicesClass IITerminated25 Dec 2019

Context

88%of 2019 device events are Class II968 events that year
4product records in this eventfirm median: 1
225events on this register for the firm

Other enforcement events for GE Healthcare, LLC

InitiatedProduct (first record)TypeClassRecords
27 Feb 2026SIGNA Premier systemsDeviceClass II1
15 Jul 2022(1) GE Centricity Universal Viewer 6.0, Model 2088026-9XX; (2) GE Centricity Universal Viewer 7.0, Model 5826659-0XX; (3DeviceClass II1
24 Jun 2022SIGNA Premier magnetic resonance scanner, model 5748519.DeviceClass II1
14 Jun 2022Centricity PACS Software Version 7.0 SP0.0.4.7DeviceClass II1
8 Jun 2022GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data withinDeviceClass II1
2 Jun 2022Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High CapDeviceClass II1
1 Jun 2022GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro Anesthesia SysteDeviceClass I1
27 May 2022GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, PartDeviceClass II1
10 May 2022CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and DeviceClass II1
28 Apr 2022Centricity Universal Viewer Zero Footprint ClientDeviceClass II5

All 225 events for GE Healthcare, LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 13 Apr 2026. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.