Recall Register

Food, drugs, devices › Maquet Cardiovascular Us Sales, Llc

Device recall: Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical

FDA enforcement event 82975 · Class II · initiated 20 May 2019 · status Terminated · terminated 15 Dec 2023

DeviceVoluntary: Firm initiated2 product recordsWayne, NJ

Maquet Cardiovascular Us Sales, Llc (Wayne, NJ) initiated this device recall on 20 May 2019; FDA lists 2 product records under event 82975, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 968 device enforcement events initiated in 2019 carry the same class.

Status: Terminated, terminated 15 Dec 2023. Initiation: voluntary: firm initiated. Maquet Cardiovascular Us Sales, Llc has 40 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Maquet Cardiovascular Us Sales, Llc, Wayne, NJ.

Reason for recall

"Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process."

Distribution

"Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1777-2019Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination…4 unitsClass IITerminated19 Jun 2019
Z-1778-2019Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.40 unitsClass IITerminated19 Jun 2019

Context

88%of 2019 device events are Class II968 events that year
2product records in this eventfirm median: 1
40events on this register for the firm

Other enforcement events for Maquet Cardiovascular Us Sales, Llc

InitiatedProduct (first record)TypeClassRecords
20 Nov 2020HLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 - Product UsagDeviceClass II2
30 Sep 2020Blood Monitoring Unit (BMU 40), model no. 701040852 - Product Usage: is to monitor blood parameters during cardiopulmonaDeviceClass II1
24 Sep 2020Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during anDeviceClass II2
7 Aug 2020Maquet/Getinge Rotaflow System Console US (115 V / 60 Hz) - Product Usage: Instructions for Use - Centrifugal Pump SysteDeviceClass III1
18 Jun 2020Holder HKHZ 19 for use with Quadrox-i and Quadrox-iD Neonatal / Pediatric Oxygenators, Model Number: 701047495 - ProductDeviceClass II1
20 Feb 2020Avalon Elite Vascular Access Kit, Model No. 70106.3540, UDI 04037691831039DeviceClass II1
10 Feb 2020Getinge 88-Series-washer disinfectors intended for cleaning, disinfection and drying Model Number: 88-5 Part Number: 88-DeviceClass II1
16 Jan 2020Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as CentraDeviceClass III2
3 Dec 2019Maquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component of the ROTAFLOW SDeviceClass II1
11 Jul 2019Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794DeviceClass II1

All 40 events for Maquet Cardiovascular Us Sales, Llc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 15 Dec 2023. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.