Food, drugs, devices › CooperSurgical, Inc.
Device recall: LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combine
CooperSurgical, Inc. (Trumbull, CT) initiated this device recall on 23 Apr 2019; FDA lists 1 product record under event 82821, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 968 device enforcement events initiated in 2019 carry the same class.
Status: Terminated, terminated 24 Apr 2020. Initiation: voluntary: firm initiated. CooperSurgical, Inc. has 57 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
CooperSurgical, Inc., Trumbull, CT.
Reason for recall
"2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses."
Distribution
"US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1840-2019 | LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure… | 19 | Class II | Terminated | 26 Jun 2019 |
Context
Other enforcement events for CooperSurgical, Inc.
All 57 events for CooperSurgical, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI."
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 24 Apr 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.