Device recall: Instrument Arm Drape (Box of 20) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the…
Class II FDA enforcement event 81244. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a breach in the sterility of the product.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Ecolab Inc started this device recall on 14 Sep 2018. The firm address in the FDA record is in Saint Paul, MN. FDA put the recall in Class II. FDA event 81244 has 4 product records. The first FDA enforcement report date is 31 Oct 2018. The record gives the status Terminated, with the termination date 12 Feb 2021. 113 of the 1,074 device events that started in 2018 have more product records. The median device event of 2018 has 1 product record. The openFDA file has 1,074 device events that started in 2018 and 1,091 that started in 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide - US Nationwide Distribution in the states of AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH,…
Recalling firm
Ecolab Inc, Saint Paul, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
4 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0231-2019 | Instrument Arm Drape (Box of 20) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system Distribute by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or nonsurgical… | 344 cases (6,880 eaches) | Class II | Terminated | 31 Oct 2018 |
| Z-0232-2019 | Camera Arm Drape (Box of 20) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or nonsurgical… | 88 cases (1,760 eaches) | Class II | Terminated | 31 Oct 2018 |
| Z-0233-2019 | Disposable Accessory Kit, 3-Arm (Box of 5) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or… | 4,861 cases (24,305 eaches) | Class II | Terminated | 31 Oct 2018 |
| Z-0234-2019 | Disposable Accessory Kit, 4-Arm (Box of 5) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or… | 9,364 cases (46,820 eaches) | Class II | Terminated | 31 Oct 2018 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2018
- 1,074
- Share of these events in Class II
- 92%
- Product records in this event
- 4Median of the firm: 1
- Events of the firm on the register
- 14
Other events of the firm
Official channels
- FDA enforcement report Z-0231-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 81244. https://recallregister.com/fda/ii/81244/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide - US Nationwide Distribution in the states of AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH,…
- What quantity of product is in FDA event 81244?
- The event has 4 product records. The first record gives this quantity:
344 cases (6,880 eaches)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2018 have the same class?
- 92% of the 1,074 device events that started in 2018 are in Class II.