Recall Register

Food, drugs, devices › GE Medical Systems, LLC

Device recall: GE Healthcare MRI Systems, Software Version 11.x, sold under the following product names: 1.5T CXK4 LCC MAGNET,1.5T EXCI

FDA enforcement event 80323 · Class II · initiated 7 Jun 2018 · status Terminated · terminated 18 Jun 2020

DeviceVoluntary: Firm initiated3 product recordsWaukesha, WI

FDA enforcement event 80323 is a device recall by GE Medical Systems, LLC of Waukesha, WI, initiated 7 Jun 2018 and classified Class II, with 3 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 92% of the 1,074 device enforcement events initiated in 2018 carry the same class.

Status: Terminated, terminated 18 Jun 2020. Initiation: voluntary: firm initiated. GE Medical Systems, LLC has 42 enforcement events on this register, with a median of 3 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

GE Medical Systems, LLC, Waukesha, WI.

Reason for recall

"The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch."

Distribution

"Worldwide Distribution: US (nationwide) including states of:: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0110-2019GE Healthcare MRI Systems, Software Version 11.x, sold under the following product names: 1.5T CXK4 LCC MAGNET,1.5T EXCITE, 1.5T EXCITE E/S, 1.5T EXCITE GS, 1.5T EXCITE HYBRID, 1.5T EXCITE TWIN, 1.5T EXCITE UPGRAD, 1.5T HD ES/HS TR W/EXCITE, 1.5T MR MOBILE (SERIAL # R350), 1.5T SIGNA ECHOSPEED, 1.5T SIGNA EXCITE,…5,200 total unitsClass IITerminated24 Oct 2018
Z-0111-2019GE Healthcare MRI Systems, Software Version 12.x, sold under the following product names: SIGNA MR/I ECHOSPD W/ EXCITE, 1.5 HISPEED HD 8-CH 400-AP, MR GOLDSEAL 1.5T HD 8CH USED, UPGRADE 1.5 EXC-II TO EXC-III HD, SIGNA MR/I HISPD W/ EXCITE, 1.5 ECHOSPEED HD 16-CH 800-AP, 1.5 HISPEED HD 4-CH 200-AP, 1.5 TWINSPEED HD…5,200 total unitsClass IITerminated24 Oct 2018
Z-0112-2019GE Healthcare MRI Systems, Software Version 15.x or HD 16 sold under the following product names: *ER*15T HDXT ECHOSPEED 16 CHANNEL, *OC*1.5T HDX ECHOSPEED 8 CHANNEL, *SC*1.5T HD TO HDX UPGRADE, *TR*15T HDXT 16CH, *TR*3.0T HDX TWINSPEED 16 CHANNEL, 1,5T HDXT, 1.5 HD TO HDX UPGRADE, 1.5 HDX, 1.5 HDX ECHOSPEED 16 CH,…5,200 total unitsClass IITerminated24 Oct 2018

Context

92%of 2018 device events are Class II1,074 events that year
3product records in this eventfirm median: 3
42events on this register for the firm

Other enforcement events for GE Medical Systems, LLC

InitiatedProduct (first record)TypeClassRecords
5 Jun 2026Optima XR200amx, Mobile Digital-Ready Radiographic SystemDeviceClass II3
1 May 2026Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All ModelsDeviceClass II3
20 Apr 2026Allia IGS 3 Pulse angiographic X-ray systemDeviceClass II5
26 Mar 2026Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300DeviceClass II7
3 Feb 2026Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E8DeviceClass II1
30 Jan 2026Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including DeviceClass II3
24 Oct 2025Senographe Pristina, Pristina Serena, and Pristina Serena 3D. for screening and diagnostic mammographyDeviceClass II1
18 Sep 2025LOGIQ P10 series with software version R4.5.7 Model Number 5877534DeviceClass II3
22 Aug 2025MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model NuDeviceClass II1
2 Jul 2025GE HealthCare OEC 9800.DeviceClass II2

All 42 events for GE Medical Systems, LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution: US (nationwide) including states of:: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 18 Jun 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.