Recall Register

Food, drugs, devices › Fresenius Medical Care Renal Therapies Group, LLC

Device recall: 2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

FDA enforcement event 79239 · Class II · initiated 5 Oct 2017 · status Terminated · terminated 28 Jul 2021

DeviceVoluntary: Firm initiated1 product recordWaltham, MA

FDA enforcement event 79239 is a device recall by Fresenius Medical Care Renal Therapies Group, LLC of Waltham, MA, initiated 5 Oct 2017 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.

Status: Terminated, terminated 28 Jul 2021. Initiation: voluntary: firm initiated. Fresenius Medical Care Renal Therapies Group, LLC has 22 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Fresenius Medical Care Renal Therapies Group, LLC, Waltham, MA.

Reason for recall

"While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also…"

Distribution

"US Nationwide Distribution"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0829-20182008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.15 machinesClass IITerminated7 Mar 2018

Context

93%of 2017 device events are Class II1,091 events that year
1product record in this eventfirm median: 1
22events on this register for the firm

Other enforcement events for Fresenius Medical Care Renal Therapies Group, LLC

InitiatedProduct (first record)TypeClassRecords
10 Aug 20265008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6DeviceClass II4
14 Jul 2026GRANUFLO Dry Acid Concentrate: Product Numbers: 0FD2251-3B; 0FD3251-3B; Dry Acid Concentrate for Bicarbonate HemodialysiDeviceClass II3
30 Jun 2026Fresenius Medical Care, 2008T BlueStar Hemodialysis Machine, Model Numbers: 191126, 191130.DeviceClass II1
29 May 2019Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjDeviceClass II1
3 Apr 20192008T BlueStar Machine Software v.2.72-Hemodialysis machine Model Number: 191126 - Product Usage: The 2008T hemodialysisDeviceClass III1
5 Nov 2018Optiflux F160NR Capillary High Flux Dialyzer Single Use Only, Product Code 0500316EDeviceClass II1
9 Oct 2018Fresenius Granuflo Dry Acid Concentrate- 2K 2.0 Ca 1 Mg 16.5 GAL Catalog Number: OFD2201-3B for hemodialysisDeviceClass III3
13 Jul 20182008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part NumbDeviceClass II2
2 May 2018NaturaLyte¿ Liquid Acid Concentrate for Bicarbonate Dialysis, 3.43 Liters, Part Number: 08-3231-1 Product Usage: This coDeviceClass II1
24 Jan 2018Liberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chroDeviceClass II1

All 22 events for Fresenius Medical Care Renal Therapies Group, LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also…"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide Distribution"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 28 Jul 2021. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.