Recall Register

Food, drugs, devices

Device recall: Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted into a shel

FDA enforcement event 78824 · Class II · initiated 14 Dec 2017 · status Terminated · terminated 14 Dec 2018

DeviceVoluntary: Firm initiated2 product recordsCampbell, CA

Kalila Medical (Campbell, CA) initiated this device recall on 14 Dec 2017; FDA lists 2 product records under event 78824, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.

Status: Terminated, terminated 14 Dec 2018. Initiation: voluntary: firm initiated. This is the only enforcement event on this register for Kalila Medical.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Kalila Medical, Campbell, CA.

Reason for recall

"Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process."

Distribution

"Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0563-2018Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.99 sheathsClass IITerminated21 Feb 2018
Z-0564-2018Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.261 sheathsClass IITerminated21 Feb 2018

Context

93%of 2017 device events are Class II1,091 events that year
2product records in this event
1event on this register for the firm

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 14 Dec 2018. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.