Food, drugs, devices › Med Tec Inc
Device recall: CIVCO Arm Support, REF 106015, distributed as: (a) Rigid Arm Support, Standard, Set Red (PN: 106015) (b) Rigid Adj. Arm
Med Tec Inc (Orange City, IA) initiated this device recall on 10 Aug 2017; FDA lists 2 product records under event 77874, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 19 Mar 2024. Initiation: voluntary: firm initiated. Med Tec Inc has 7 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Med Tec Inc, Orange City, IA.
Reason for recall
"Rigid Arm Supports (PN: 106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm."
Distribution
"Worldwide Distribution - US Distribution to the states of : MI, IL and TX., and to the countries of : Greece, Australia, Czech Republic, Hungary, Norway, Brazil, Italy, Switzerland ,UK, Austria,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-3082-2017 | CIVCO Arm Support, REF 106015, distributed as: (a) Rigid Arm Support, Standard, Set Red (PN: 106015) (b) Rigid Adj. Arm Support, Set (PN: 106020) (c) PSR2-CR-DLR-2P Posirest-2 (PN: 106521) (d) PB2-BLR Posiboard-2 (PN: 109030) The Rigid Arm Supports are utilized on the Posibboard2 and Posirest assemblies, which are… | 101 units | Class II | Terminated | 6 Sep 2017 |
| Z-3083-2017 | CIVCO Arm Support, REF 106047, distributed as: (a) Rigid Arm Support, Large, Set, Red (PN: 106047) (b) PB2-BDLRL (PN: 109040) (c) Rigid Adj. Arm Support, Large, Set (PN: 106048) (d) PSR2-CF-DLRL-2P Posirest-2 (PN: 106522) The Rigid Arm Supports are utilized on the Posibboard2 and Posirest assemblies, which are… | 30 units | Class II | Terminated | 6 Sep 2017 |
Context
Other enforcement events for Med Tec Inc
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 12 May 2020 | Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioni | Device | Class II | 1 |
| 15 Apr 2019 | CIVCO Solstice(TM) SRS Immobilization System, Part Numbers: a) CHS03, b) CHS04 - Product Usage: The Tilting Head Fixatio | Device | Class II | 1 |
| 8 Nov 2017 | Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XS | Device | Class II | 1 |
| 31 Mar 2016 | Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1. | Device | Class II | 1 |
| 19 Aug 2015 | CIVCO Shoulder Retractors, Reusable non-sterile hand grip, REF 302070, packaged 1 set per box, RX. For use in patient po | Device | Class II | 1 |
| 11 Dec 2014 | Civco, MTSR02 Shoulder Retractor with Sherpa Wrist Loops Kit with Base, Non-sterile re-usable wool wrist strap, REF: 20S | Device | Class II | 2 |
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Rigid Arm Supports (PN: 106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - US Distribution to the states of : MI, IL and TX., and to the countries of : Greece, Australia, Czech Republic, Hungary, Norway, Brazil, Italy, Switzerland ,UK, Austria,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 19 Mar 2024. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance