Food, drugs, devices › Carestream Health Inc
Device recall: DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China onl
Carestream Health Inc (Rochester, NY) initiated this device recall on 25 May 2017; FDA lists 1 product record under event 77250, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 13 Jun 2018. Initiation: voluntary: firm initiated. Carestream Health Inc has 30 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Carestream Health Inc, Rochester, NY.
Reason for recall
"Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly."
Distribution
"Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2651-2017 | DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures | Domestic: 1449 units; Foreign: 1416 units | Class II | Terminated | 5 Jul 2017 |
Context
Other enforcement events for Carestream Health Inc
All 30 events for Carestream Health Inc
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 13 Jun 2018. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.