Device recall: UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.
FDA enforcement event 77145 is a device recall by Beckman Coulter Inc. of Brea, CA, initiated 24 Apr 2017 and classified Class II, with 2 product records.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 22 May 2018. Initiation: voluntary: firm initiated. Beckman Coulter Inc. has 216 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Beckman Coulter Inc., Brea, CA.
Reason for recall
"There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result."
Distribution
"Worldwide distribution: US nationwide, Algeria, Andorra, Austria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Cote…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2157-2017 | UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322. | 4880 units total (2119 units in US) | Class II | Terminated | 31 May 2017 |
| Z-2158-2017 | UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858. | 1408 units total (998 units in US) | Class II | Terminated | 31 May 2017 |
Context
Other enforcement events for Beckman Coulter Inc.
All 216 events for Beckman Coulter Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide distribution: US nationwide, Algeria, Andorra, Austria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Cote…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 22 May 2018. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.