Food, drugs, devices › Encore Medical, Lp
Device recall: LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a
FDA enforcement event 76209 is a device recall by Encore Medical, Lp of Austin, TX, initiated 10 Jan 2017 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 31 Mar 2017. Initiation: voluntary: firm initiated. Encore Medical, Lp has 61 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Encore Medical, Lp, Austin, TX.
Reason for recall
"Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization."
Distribution
"US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1191-2017 | LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both… | 13 devices | Class II | Terminated | 22 Feb 2017 |
Context
Other enforcement events for Encore Medical, Lp
All 61 events for Encore Medical, Lp
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 31 Mar 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.