Device recall: Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.
The Binding Site Group, Ltd. (Birmingham) initiated this device recall on 5 Jul 2016; FDA lists 1 product record under event 76048, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,232 device enforcement events initiated in 2016 carry the same class.
Status: Terminated, terminated 3 Jul 2017. Initiation: voluntary: firm initiated. The Binding Site Group, Ltd. has 23 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
The Binding Site Group, Ltd., Birmingham, United Kingdom.
Reason for recall
"Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards."
Distribution
"U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1043-2017 | Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit. | 809 kits | Class II | Terminated | 25 Jan 2017 |
Context
Other enforcement events for The Binding Site Group, Ltd.
All 23 events for The Binding Site Group, Ltd.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium."
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 3 Jul 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.