Device recall: Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement o
FDA enforcement event 75709 is a device recall by Roche Diagnostics Operations, Inc. of Indianapolis, IN, initiated 3 Sep 2015 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.
Status: Terminated, terminated 5 Jun 2017. Initiation: voluntary: firm initiated. Roche Diagnostics Operations, Inc. has 69 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Roche Diagnostics Operations, Inc., Indianapolis, IN.
Reason for recall
"under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously,…"
Distribution
"Nationwide Distribution including CA, LA, NE, PR, SC, TX, VA, and WV"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1077-2017 | Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes Nat, K+, iCaWt (ISE), hematocrit (THct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COT~b, MetHb), and oxygen saturation… | 30 | Class II | Terminated | 1 Feb 2017 |
Context
Other enforcement events for Roche Diagnostics Operations, Inc.
All 69 events for Roche Diagnostics Operations, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously,…"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Nationwide Distribution including CA, LA, NE, PR, SC, TX, VA, and WV"
Is the recall still ongoing?