Device recall: The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of…

Class II FDA enforcement event 73033. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Horiba Instruments Inc started this device recall on 11 Jan 2016. The firm address in the FDA record is in Irvine, CA. FDA put the recall in Class II. FDA event 73033 has 6 product records. The first FDA enforcement report date is 10 Feb 2016. The record gives the status Terminated, with the termination date 7 Jun 2016. 71 of the 1,232 device events that started in 2016 have more product records. The median device event of 2016 has 1 product record. The openFDA file has 1,232 device events that started in 2016 and 1,153 that started in 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.

Recalling firm

Horiba Instruments Inc, Irvine, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 73033: product records
Recall numberProductQuantityClassStatusReport date
Z-0738-2016The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of cholesterol in human serum and plasma based on an enzymatic photometric test (Trinder’s reaction). The assay is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.216Class IITerminated10 Feb 2016
Z-0739-2016ABX PENTRA Enzymatic Creatinine CP ref. A11A01907 reagent is intended for the quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine based on an enzymatic method of using a multi-step approach ending with a photometric end-point reaction.216Class IITerminated10 Feb 2016
Z-0740-2016ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.216Class IITerminated10 Feb 2016
Z-0741-2016ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).216Class IITerminated10 Feb 2016
Z-0742-2016ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determination of triglycerides in human serum and plasma based on an enzymatic colorimetric test. It is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.216Class IITerminated10 Feb 2016
Z-0743-2016ABX PENTRA Uric Acid CP ref. A11A01670 is an in vitro diagnostic assay for the quantitative determination of uric acid in human serum, plasma and urine based on the enzymatic determination of uric acid using a chromogenic system in the presence of peroxidase and uricase (Trinder method).216Class IITerminated10 Feb 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2016
1,232
Share of these events in Class II
91%
Product records in this event
6Median of the firm: 2
Events of the firm on the register
3

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 73033. https://recallregister.com/fda/ii/73033/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
What quantity of product is in FDA event 73033?
The event has 6 product records. The first record gives this quantity: 216
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2016 have the same class?
91% of the 1,232 device events that started in 2016 are in Class II.