Device recall: Bond Polymer Refine Red Detection DS9390 kits. The common name of the device is detection system for in vitro diagnostic
Leica Microsystems, Inc. (Buffalo Grove, IL) initiated this device recall on 27 Nov 2015; FDA lists 1 product record under event 72802, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.
Status: Terminated, terminated 5 Jan 2017. Initiation: voluntary: firm initiated. Leica Microsystems, Inc. has 33 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Leica Microsystems, Inc., Buffalo Grove, IL.
Reason for recall
"The Firm informed the importer/distributor, Leica Microsystems, a manufacturing error occurred during the filling of some containers of the Bond Polymer Refine Red Detection DS9390 kits, which may result in inadequate staining."
Distribution
"Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Argentina, Australia, Austria, Belgium, Chile, Denmark, France, Germany, Greece, Hong Kong, Italy, Japan ,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0808-2016 | Bond Polymer Refine Red Detection DS9390 kits. The common name of the device is detection system for in vitro diagnostic use. The Intended Use states that Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphatase (AP)-linker antibody conjugate system for the detection of tissue-bound mouse… | 3745 | Class II | Terminated | 24 Feb 2016 |
Context
Other enforcement events for Leica Microsystems, Inc.
All 33 events for Leica Microsystems, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The Firm informed the importer/distributor, Leica Microsystems, a manufacturing error occurred during the filling of some containers of the Bond Polymer Refine Red Detection DS9390 kits, which may result in inadequate staining."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Argentina, Australia, Austria, Belgium, Chile, Denmark, France, Germany, Greece, Hong Kong, Italy, Japan ,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 5 Jan 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.