Recall Register

Food, drugs, devices › Intuitive Surgical, Inc.

Device recall: 0 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General and Plastic Surgery: The da Vinci

FDA enforcement event 72776 · Class II · initiated 23 Nov 2015 · status Terminated · terminated 5 Apr 2016

DeviceVoluntary: Firm initiated2 product recordsSunnyvale, CA

Intuitive Surgical, Inc. (Sunnyvale, CA) initiated this device recall on 23 Nov 2015; FDA lists 2 product records under event 72776, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.

Status: Terminated, terminated 5 Apr 2016. Initiation: voluntary: firm initiated. Intuitive Surgical, Inc. has 105 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Intuitive Surgical, Inc., Sunnyvale, CA.

Reason for recall

"This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources."

Distribution

"Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0652-20160 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision Cart…2624 total all modelsClass IITerminated20 Jan 2016
Z-0653-201630 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on the Surgeon Console 3D viewer and Vision…2624 total all modelsClass IITerminated20 Jan 2016

Context

91%of 2015 device events are Class II1,153 events that year
2product records in this eventfirm median: 1
105events on this register for the firm

Other enforcement events for Intuitive Surgical, Inc.

InitiatedProduct (first record)TypeClassRecords
24 Apr 2026Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product DeDeviceClass II1
22 Apr 2026Da Vinci ASSY, PSS, SP1098, Part Number: 380601DeviceClass II1
2 Apr 2026Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200DeviceClass II1
11 Mar 2026Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06DeviceClass I1
30 Dec 2025da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45DeviceClass II1
15 Dec 2025da Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073 (2.7-4cm,SP1098, Box) & 430075 (2.7-7cm,SP1098,DeviceClass II1
9 Dec 2025Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/DeviceClass II2
9 Dec 2025Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved ScissorDeviceClass II4
4 Dec 2025Brand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog Number: 490127 Software Version: N/ADeviceClass II1
17 Nov 2025Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally invasive surgery.DeviceClass II1

All 105 events for Intuitive Surgical, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Monaco,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 5 Apr 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.