Food, drugs, devices › Beckman Coulter Inc.
Device recall: Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobu
Beckman Coulter Inc. (Brea, CA) initiated this device recall on 28 Jul 2015; FDA lists 1 product record under event 71878, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.
Status: Terminated, terminated 13 Apr 2016. Initiation: voluntary: firm initiated. Beckman Coulter Inc. has 216 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Beckman Coulter Inc., Brea, CA.
Reason for recall
"Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL."
Distribution
"Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2733-2015 | Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems. | 3,035 (U.S.) and 7,123 (International) | Class II | Terminated | 9 Sep 2015 |
Context
Other enforcement events for Beckman Coulter Inc.
All 216 events for Beckman Coulter Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 13 Apr 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.