Recall Register

Food, drugs, devices › RAYSEARCH LABORATORIES AB

Device recall: RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system d

FDA enforcement event 71289 · Class II · initiated 8 May 2015 · status Terminated · terminated 11 Apr 2017

DeviceVoluntary: Firm initiated1 product record

RAYSEARCH LABORATORIES AB (Stockholm) initiated this device recall on 8 May 2015; FDA lists 1 product record under event 71289, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.

Status: Terminated, terminated 11 Apr 2017. Initiation: voluntary: firm initiated. RAYSEARCH LABORATORIES AB has 57 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

RAYSEARCH LABORATORIES AB, Stockholm, Sweden.

Reason for recall

"RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing."

Distribution

"US Nationwide Distribution"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1679-2015RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed…Domestic: 539 unitsClass IITerminated3 Jun 2015

Context

91%of 2015 device events are Class II1,153 events that year
1product record in this eventfirm median: 1
57events on this register for the firm

Other enforcement events for RAYSEARCH LABORATORIES AB

InitiatedProduct (first record)TypeClassRecords
20 May 2026RayCare 2024A SP1. Oncology Information Systems.DeviceClass II2
28 Nov 2025RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment PlDeviceClass II9
8 Aug 2025RayStation with the following product descriptions: 1. RayStation 7, Software Version: 7.0.0.19. 2. RayStation 8A, SoftwDeviceClass II9
16 Jan 2025Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.DeviceClass II11
28 Mar 2024RayStation 7.0.0.19. Radiation Therapy Treatment Planning System.DeviceClass II12
23 Feb 2024RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including somDeviceClass II1
12 Sep 2023RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as wDeviceClass II1
17 Jul 2023RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to suppDeviceClass II1
17 Feb 2023Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10DeviceClass II1
10 Jun 2022RayStation 11BDeviceClass II1

All 57 events for RAYSEARCH LABORATORIES AB

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide Distribution"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 11 Apr 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.