Recall Register

Food, drugs, devices › GE Healthcare

Device recall: D-fend, (p/n 876446-HEL), and D-fend+ (p/n 881319-HEL) water trap. Used with Compact Airway Modules E/M-C(Ai)O(V)(X) and

FDA enforcement event 71044 · Class II · initiated 24 Apr 2015 · status Terminated · terminated 7 Sep 2016

DeviceVoluntary: Firm initiated1 product recordWaukesha, WI

FDA enforcement event 71044 is a device recall by GE Healthcare of Waukesha, WI, initiated 24 Apr 2015 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,153 device enforcement events initiated in 2015 carry the same class.

Status: Terminated, terminated 7 Sep 2016. Initiation: voluntary: firm initiated. GE Healthcare has 44 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

GE Healthcare, Waukesha, WI.

Reason for recall

"A potential safety issue due to loss of gas monitoring associated with occlusion of specific D-Fend / D-Fend+ water traps used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. This water trap occlusion could cause a loss of respiratory airway gas monitoring."

Distribution

"Worldwide Distribution - US Nationwide including Washington DC, Puerto Rico. and the countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1680-2015D-fend, (p/n 876446-HEL), and D-fend+ (p/n 881319-HEL) water trap. Used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. Product Usage: Water traps are intended to protect the GE monitor from humidity, secretions and bacterial contamination when monitoring respiratory parameters. D-fend and…5,018 (US = 2,948 / OUS = 2,070)Class IITerminated3 Jun 2015

Context

91%of 2015 device events are Class II1,153 events that year
1product record in this eventfirm median: 1
44events on this register for the firm

Other enforcement events for GE Healthcare

InitiatedProduct (first record)TypeClassRecords
31 Jul 2026Centricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving, and sending electronic medical DeviceClass II6
16 Jan 2026GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; SDeviceClass II1
19 Feb 2025GE Healthcare Centricity Universal Viewer (UV), Model Numbers: 1) 2066908-077; 2) 2066908-136; 3) 2066908-150; 4) 208802DeviceClass II5
3 Oct 2024GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 20895DeviceClass II1
29 Aug 2024GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a)¿¿¿¿¿ 2089507-120; b)¿¿¿¿DeviceClass II1
5 Aug 2024GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.DeviceClass II1
23 Jul 2024GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 21103DeviceClass II1
14 Sep 2023Centricity PACS-IW, software versions V3.7.x through 3.7.3.9 SP3; Radiological Image Processing SystemDeviceClass II3
8 Sep 2023Centricity Universal Viewer Workflow Manager, Image processing radiological systemDeviceClass II2
28 Mar 2023Centricity PACS-IW with Universal Viewer, Model/Catalog Numbers: a) 2066908-061, b) 2066908-077, c) 2066908-086, d) 2066DeviceClass II3

All 44 events for GE Healthcare

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "A potential safety issue due to loss of gas monitoring associated with occlusion of specific D-Fend / D-Fend+ water traps used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. This water trap occlusion could cause a loss of respiratory airway gas monitoring."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - US Nationwide including Washington DC, Puerto Rico. and the countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 7 Sep 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.