Device recall: LIFEPAK 1000 defibrillator. Affected Part Numbers: 320371500XXX The LP1000 (in AED mode) is indicated for use on patient
Physio-Control, Inc. (Redmond, WA) initiated this device recall on 9 May 2014; FDA lists 1 product record under event 67963, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,176 device enforcement events initiated in 2014 carry the same class.
Status: Terminated, terminated 29 May 2018. Initiation: voluntary: firm initiated. Physio-Control, Inc. has 27 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Physio-Control, Inc., Redmond, WA.
Reason for recall
"Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated. A software malfunction in the LIFEPAK 100"
Distribution
"Worldwide Distribution-USA (nationwide) and the countries of ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BARBADOS, BELGIUM, BOLIVIA, BOSNIA AND HERZOGOVENIA, BRAZIL, BRUNEI…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1660-2014 | LIFEPAK 1000 defibrillator. Affected Part Numbers: 320371500XXX The LP1000 (in AED mode) is indicated for use on patients in cardiac arrest. | 96,611 units total (40,974 in the US and 55,637 outside the US) | Class II | Terminated | 4 Jun 2014 |
Context
Other enforcement events for Physio-Control, Inc.
All 27 events for Physio-Control, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated. A software malfunction in the LIFEPAK 100"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution-USA (nationwide) and the countries of ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BARBADOS, BELGIUM, BOLIVIA, BOSNIA AND HERZOGOVENIA, BRAZIL, BRUNEI…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 29 May 2018. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.