Food, drugs, devices › Maquet Cardiovascular Us Sales, Llc
Device recall: MAQUET PLEGIOX Cardioplegia Heat Exchanger The PLEGIOX Cardioplegia Heat Exchanger is used as part of a cardioplegia sys
FDA enforcement event 67515 is a device recall by Maquet Cardiovascular Us Sales, Llc of Wayne, NJ, initiated 31 Jan 2014 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 91% of the 1,176 device enforcement events initiated in 2014 carry the same class.
Status: Terminated, terminated 24 Apr 2015. Initiation: voluntary: firm initiated. Maquet Cardiovascular Us Sales, Llc has 40 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Maquet Cardiovascular Us Sales, Llc, Wayne, NJ.
Reason for recall
"The MAQUET post-marketing vigilance program has identified a discrepancy that the desired cardioplegia temperature may not be obtained when the PLEGIOX Cardioplegia Heat Exchanger (PLEGIOX) is used in conjunction with a heater/cooler unit."
Distribution
"Worldwide Distribution - USA (nationwide) and the countries of Brazil, Germany and Turkey"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1165-2014 | MAQUET PLEGIOX Cardioplegia Heat Exchanger The PLEGIOX Cardioplegia Heat Exchanger is used as part of a cardioplegia system to set and maintain temperature for given flow rates and within the given temperature range of blood cardioplegic and crystalloid cardioplegic solutions during extracorporeal circulation. The… | 1,424 units | Class II | Terminated | 19 Mar 2014 |
Context
Other enforcement events for Maquet Cardiovascular Us Sales, Llc
All 40 events for Maquet Cardiovascular Us Sales, Llc
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The MAQUET post-marketing vigilance program has identified a discrepancy that the desired cardioplegia temperature may not be obtained when the PLEGIOX Cardioplegia Heat Exchanger (PLEGIOX) is used in conjunction with a heater/cooler unit."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - USA (nationwide) and the countries of Brazil, Germany and Turkey"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 24 Apr 2015. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.