Food, drugs, devices › Beckman Coulter Inc.
Device recall: UniCel DxH 800 Coulter Cellular Analysis System, Part No. 629029, B24465, and B24802. Used for in vitro diagnostic use i
FDA enforcement event 67058 is a device recall by Beckman Coulter Inc. of Brea, CA, initiated 8 Nov 2013 and classified Class II, with 4 product records.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 90% of the 1,236 device enforcement events initiated in 2013 carry the same class.
Status: Terminated, terminated 25 Dec 2014. Initiation: voluntary: firm initiated. Beckman Coulter Inc. has 216 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Beckman Coulter Inc., Brea, CA.
Reason for recall
"Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell differential and the reticulocyte."
Distribution
"Worldwide Distribution -- US, Canada, Australia, China, Belgium, Brunei Darssalam, Colombia, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Italy,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1828-2014 | UniCel DxH 800 Coulter Cellular Analysis System, Part No. 629029, B24465, and B24802. Used for in vitro diagnostic use in screening patient populations found in clinical laboratories. | 2,874 units total (1696 units in US) | Class II | Terminated | 2 Jul 2014 |
| Z-1829-2014 | UniCel DxH 600 Coulter Cellular Analysis System, Part No. B23858. Used for in vitro diagnostic use in screening patient populations found in clinical laboratories. | 164 units total (106 units in US) | Class II | Terminated | 2 Jul 2014 |
| Z-1830-2014 | Coulter DxH Cleaner 5L, Part No. 628022. For use on UniCel DxH Systems for components that come in contact with blood samples. | 768 units total (6 units in US) | Class II | Terminated | 2 Jul 2014 |
| Z-1831-2014 | Coulter DxH Cleaner 10L, Part No. 628023. For use on UniCel DxH Systems for components that come in contact with blood samples. | 12,924 units total (9,733 units in US) | Class II | Terminated | 2 Jul 2014 |
Context
Other enforcement events for Beckman Coulter Inc.
All 216 events for Beckman Coulter Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell differential and the reticulocyte."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution -- US, Canada, Australia, China, Belgium, Brunei Darssalam, Colombia, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Italy,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 25 Dec 2014. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.