Device recall: Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to…
Class II FDA enforcement event 65406. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during use of the Panta Nail Support Device, Integra has decided to issue this voluntary medical device correction. The misalignment can result in the inability to properly insert the screws into the Panta Nail and create the potential need for revision surgery.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Integra LifeSciences Corporation started this device recall on 9 May 2013. The firm address in the FDA record is in Plainsboro, NJ. FDA put the recall in Class II. FDA event 65406 has 1 product record. The first FDA enforcement report date is 24 Jul 2013. The record gives the status Terminated, with the termination date 7 Nov 2013. 277 of the 1,236 device events that started in 2013 have more product records. The median device event of 2013 has 1 product record. The openFDA file has 1,236 device events that started in 2013 and 823 that started in 2012.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution-USA (nationwide) and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Israel, Italy,…
Recalling firm
Integra LifeSciences Corporation, Plainsboro, NJ.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1736-2013 | Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. | 692 | Class II | Terminated | 24 Jul 2013 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2013
- 1,236
- Share of these events in Class II
- 90%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 7
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 11 Mar 2016 | Integra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra Dermal Regeneration Template, | Device | Class II | 1 |
| 19 Dec 2014 | The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provides the abi | Device | Class II | 1 |
| 28 Mar 2014 | Integra Newdeal SURFIX Fixed Angle Locking Screw 3.5mm x 22mm (catalog number 285322S) and 3.5mm x 30mm (catalog number | Device | Class II | 1 |
| 28 Mar 2013 | Dural Graft Matrix DuraGen Dural Graft Matrix is an absorbable implant for repair of dural defects. DuraGen matrix is an | Device | Class II | 21 |
| 5 Nov 2012 | MAYFIELD Composite Series Base Unit, Model A3101, a Swivel Adaptor (A3018) is an integral component of the Base Unit, Mo | Device | Class II | 2 |
| 9 Aug 2012 | Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly with Intubation Mounts The CRW | Device | Class II | 2 |
Official channels
- FDA enforcement report Z-1736-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 65406. https://recallregister.com/fda/ii/65406/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution-USA (nationwide) and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Israel, Italy,…
- What quantity of product is in FDA event 65406?
- The FDA record gives this quantity:
692
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2013 have the same class?
- 90% of the 1,236 device events that started in 2013 are in Class II.