Device recall: The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The…

Class II FDA enforcement event 65214. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system p

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Sysmex America, Inc. started this device recall on 31 Mar 2013. The firm address in the FDA record is in Mundelein, IL. FDA put the recall in Class II. FDA event 65214 has 1 product record. The first FDA enforcement report date is 19 Feb 2014. The record gives the status Terminated, with the termination date 15 Aug 2016. 277 of the 1,236 device events that started in 2013 have more product records. The median device event of 2013 has 1 product record. The openFDA file has 1,236 device events that started in 2013 and 823 that started in 2012.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IS, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM,…

Recalling firm

Sysmex America, Inc., Mundelein, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 65214: product records
Recall numberProductQuantityClassStatusReport date
Z-0949-2014The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The instrument can perform screening for abnormal samples with a high degree of accuracy, so it increases automation and efficiency in laboratories.214 unitsClass IITerminated19 Feb 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
12

Other events of the firm

Other FDA enforcement events of Sysmex America, Inc.
InitiatedProduct (first record)TypeClassRecords
15 Dec 2025TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running sDeviceClass II1
6 Jan 2025Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for auDeviceClass II1
6 Jul 2020Sysmex PS-10 Sample Preparation System Catalog number:BQ716341 - Product Usage: is a fully automated, user configurable,DeviceClass II1
9 Mar 2020The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smearsDeviceClass II1
21 Jan 2020Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.DeviceClass II1
27 Jul 2018Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc., Lincolnshire, IL.DeviceClass II1
8 Sep 2017Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of plateleDeviceClass II1
23 Mar 2017Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073DeviceClass II1
1 Jul 2016Fluorocell PLT reagent, Catalog #CD994563 Fluorocell PLT is used to stain platelets in a diluted blood sample for the coDeviceClass II1
15 May 2013pocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1 WakinohamaDeviceClass II1

Sysmex America, Inc.: all 12 events

Official channels

Cite this page

Recall Register. FDA enforcement event 65214. https://recallregister.com/fda/ii/65214/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IS, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM,…
What quantity of product is in FDA event 65214?
The FDA record gives this quantity: 214 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.