Device recall: CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these products is…
Class II FDA enforcement event 64589. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Trophy Sas started this device recall on 18 Sep 2012. The firm address in the FDA record is in Croissy Beaubourg. FDA put the recall in Class II. FDA event 64589 has 4 product records. The first FDA enforcement report date is 8 May 2013. The record gives the status Terminated, with the termination date 2 May 2017. 55 of the 823 device events that started in 2012 have more product records. The median device event of 2012 has 1 product record. The openFDA file has 823 device events that started in 2012 and 114 that started in 2011.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
USA Nationwide Distribution
Recalling firm
Trophy Sas, Croissy Beaubourg, France.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
4 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0999-2013 | CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures. | 167 installed in the US | Class II | Terminated | 8 May 2013 |
| Z-1000-2013 | CareStream Dental CS 2200 X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures. | 167 installed in the US | Class II | Terminated | 8 May 2013 |
| Z-1094-2013 | CareStream Dental Kodak 2100 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures. | 167 installed in the US | Class II | Terminated | 8 May 2013 |
| Z-1095-2013 | CareStream Dental Kodak 2200 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures. | 167 installed in the US | Class II | Terminated | 8 May 2013 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2012
- 823
- Share of these events in Class II
- 88%
- Product records in this event
- 4Median of the firm: 4
- Events of the firm on the register
- 2
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 21 Jul 2021 | CS9600¿isanextraoralsystemintendedtoproducetwo-dimensionalandthree-dimensionaldigital X-rayimagesofthedento-maxilofacial Event 92206 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0999-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 64589. https://recallregister.com/fda/ii/64589/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
USA Nationwide Distribution
- What quantity of product is in FDA event 64589?
- The event has 4 product records. The first record gives this quantity:
167 installed in the US
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2012 have the same class?
- 88% of the 823 device events that started in 2012 are in Class II.