Recall Register

Food, drugs, devices

Device recall: CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these products is for

FDA enforcement event 64589 · Class II · initiated 18 Sep 2012 · status Terminated · terminated 2 May 2017

DeviceVoluntary: Firm initiated4 product records

FDA enforcement event 64589 is a device recall by Trophy Sas of Croissy Beaubourg, initiated 18 Sep 2012 and classified Class II, with 4 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 823 device enforcement events initiated in 2012 carry the same class.

Status: Terminated, terminated 2 May 2017. Initiation: voluntary: firm initiated. Trophy Sas has 2 enforcement events on this register, with a median of 4 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Trophy Sas, Croissy Beaubourg, France.

Reason for recall

"There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator."

Distribution

"USA Nationwide Distribution"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0999-2013CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.167 installed in the USClass IITerminated8 May 2013
Z-1000-2013CareStream Dental CS 2200 X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.167 installed in the USClass IITerminated8 May 2013
Z-1094-2013CareStream Dental Kodak 2100 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.167 installed in the USClass IITerminated8 May 2013
Z-1095-2013CareStream Dental Kodak 2200 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.167 installed in the USClass IITerminated8 May 2013

Context

88%of 2012 device events are Class II823 events that year
4product records in this eventfirm median: 4
2events on this register for the firm

Other enforcement events for Trophy Sas

InitiatedProduct (first record)TypeClassRecords
21 Jul 2021CS¿9600¿is¿an¿extraoral¿system¿intended¿to¿produce¿two-dimensional¿and¿three-dimensional¿digital¿X-ray¿images¿of¿the¿den (92206)DeviceClass II1

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "USA Nationwide Distribution"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 2 May 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.