Device recall: Access Immunoassay System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices…

Class II FDA enforcement event 64332. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Beckman Coulter Inc. started this device recall on 12 Feb 2013. The firm address in the FDA record is in Brea, CA. FDA put the recall in Class II. FDA event 64332 has 4 product records. The first FDA enforcement report date is 10 Apr 2013. The record gives the status Terminated, with the termination date 23 Dec 2013. 78 of the 1,236 device events that started in 2013 have more product records. The median device event of 2013 has 1 product record. The openFDA file has 1,236 device events that started in 2013 and 823 that started in 2012.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy,…

Recalling firm

Beckman Coulter Inc., Brea, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 64332: product records
Recall numberProductQuantityClassStatusReport date
Z-1032-2013Access Immunoassay System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.579 unitsClass IITerminated10 Apr 2013
Z-1033-2013Access 2 Immunoassay System; P/N 107405S 2/12 The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.3500 unitsClass IITerminated10 Apr 2013
Z-1034-2013Synchron LXi 725 Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.83 unitsClass IITerminated10 Apr 2013
Z-1035-2013UniCel DxC 600i Synchron Access Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.956 unitsClass IITerminated10 Apr 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class II
90%
Product records in this event
4Median of the firm: 1
Events of the firm on the register
219

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 64332. https://recallregister.com/fda/ii/64332/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy,…
What quantity of product is in FDA event 64332?
The event has 4 product records. The first record gives this quantity: 579 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
90% of the 1,236 device events that started in 2013 are in Class II.