Device recall: Access Immunoassay System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices…
Class II FDA enforcement event 64332. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Beckman Coulter Inc. started this device recall on 12 Feb 2013. The firm address in the FDA record is in Brea, CA. FDA put the recall in Class II. FDA event 64332 has 4 product records. The first FDA enforcement report date is 10 Apr 2013. The record gives the status Terminated, with the termination date 23 Dec 2013. 78 of the 1,236 device events that started in 2013 have more product records. The median device event of 2013 has 1 product record. The openFDA file has 1,236 device events that started in 2013 and 823 that started in 2012.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy,…
Recalling firm
Beckman Coulter Inc., Brea, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
4 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1032-2013 | Access Immunoassay System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. | 579 units | Class II | Terminated | 10 Apr 2013 |
| Z-1033-2013 | Access 2 Immunoassay System; P/N 107405S 2/12 The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. | 3500 units | Class II | Terminated | 10 Apr 2013 |
| Z-1034-2013 | Synchron LXi 725 Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. | 83 units | Class II | Terminated | 10 Apr 2013 |
| Z-1035-2013 | UniCel DxC 600i Synchron Access Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. | 956 units | Class II | Terminated | 10 Apr 2013 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2013
- 1,236
- Share of these events in Class II
- 90%
- Product records in this event
- 4Median of the firm: 1
- Events of the firm on the register
- 219
Other events of the firm
Official channels
- FDA enforcement report Z-1032-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 64332. https://recallregister.com/fda/ii/64332/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy,…
- What quantity of product is in FDA event 64332?
- The event has 4 product records. The first record gives this quantity:
579 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2013 have the same class?
- 90% of the 1,236 device events that started in 2013 are in Class II.