Food, drugs, devices › Hospira Inc.
Device recall: Plum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Inf
FDA enforcement event 64091 is a device recall by Hospira Inc. of Lake Forest, IL, initiated 5 Feb 2012 and classified Class II, with 7 product records.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 823 device enforcement events initiated in 2012 carry the same class.
Status: Terminated, terminated 14 Nov 2016. Initiation: voluntary: firm initiated. Hospira Inc. has 151 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Hospira Inc., Lake Forest, IL.
Reason for recall
"Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not"
Distribution
"Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0203-2014 | Plum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care | List No. 11005: 179 units | Class II | Terminated | 20 Nov 2013 |
| Z-0204-2014 | Plum A+ Infusion Pump; List Number: 11971; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care | List No. 11971: 10,669 units | Class II | Terminated | 20 Nov 2013 |
| Z-0205-2014 | Plum A+ Infusion Pump; List Number: 11973; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care | List No. 11973: 71 units | Class II | Terminated | 20 Nov 2013 |
| Z-0206-2014 | Plum A+3 Infusion Pump System; List Number: 12348; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care | List No. 12348: 214 units | Class II | Terminated | 20 Nov 2013 |
| Z-0207-2014 | Plum A+ Infusion Pump; List Number: 12391; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care | List No. 12391: 182,970 units | Class II | Terminated | 20 Nov 2013 |
| Z-0208-2014 | Plum A+ Infusion Pump; List Number: 12618; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care | List No. 12618: 7,691 units | Class II | Terminated | 20 Nov 2013 |
| Z-0209-2014 | *** 1) Plum A+ Infusion Pump; List Number: 20678; Labeled in part: PLUM A+3; for use with Hospira MedNet; *** 2) Plum A+ Infusion Pump; List Number: 20679; Labeled in part: PLUM A+; for use with Hospira MedNet; *** 3) Plum A+ Infusion Pump with Hospira MedNet Software; List Number: 20792; Labeled in part: PLUM A+;… | 1) List No. 20678: 21,551 units; 2) List No. 20679: 7,737 units; 3) List No. 20792: 126,696 units | Class II | Terminated | 20 Nov 2013 |
Context
Other enforcement events for Hospira Inc.
All 151 events for Hospira Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 14 Nov 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.