Device recall: Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16026. Note…
Class II FDA enforcement event 64005. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay or interruption of therapy.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Hospira Inc. started this device recall on 16 Mar 2011. The firm address in the FDA record is in Lake Forest, IL. FDA put the recall in Class II. FDA event 64005 has 2 product records. The first FDA enforcement report date is 16 Nov 2016. The record gives the status Terminated, with the termination date 28 Nov 2016. 18 of the 114 device events that started in 2011 have more product records. The median device event of 2011 has 1 product record. The openFDA file has 114 device events that started in 2011 and 37 that started in 2010.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - US (nationwide) and Canada
Recalling firm
Hospira Inc., Lake Forest, IL.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0361-2017 | Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16026. Note that Hospira retired this pump June 30, 2015. The Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood, and blood… | 10,337 | Class II | Terminated | 16 Nov 2016 |
| Z-0362-2017 | Symbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16027. Note that Hospira retired this pump June 30, 2015. The Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood, and blood… | 10897 | Class II | Terminated | 16 Nov 2016 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2011
- 114
- Share of these events in Class II
- 89%
- Product records in this event
- 2Median of the firm: 1
- Events of the firm on the register
- 151
Other events of the firm
Official channels
- FDA enforcement report Z-0361-2017The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 64005. https://recallregister.com/fda/ii/64005/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - US (nationwide) and Canada
- What quantity of product is in FDA event 64005?
- The event has 2 product records. The first record gives this quantity:
10,337
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2011 have the same class?
- 89% of the 114 device events that started in 2011 are in Class II.