Device recall: KING LAD SILI CONE SINGLE USE, King Systems devices/case, ETO Sterile. Size 2 / Infant Silicone King LAD Patient Size…
Class II FDA enforcement event 63580. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
King Systems Corp. started this device recall on 2 Nov 2012. The firm address in the FDA record is in Noblesville, IN. FDA put the recall in Class II. FDA event 63580 has 10 product records. The first FDA enforcement report date is 28 Nov 2012. The record gives the status Terminated, with the termination date 28 May 2013. 19 of the 823 device events that started in 2012 have more product records. The median device event of 2012 has 1 product record. The openFDA file has 823 device events that started in 2012 and 114 that started in 2011.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
Recalling firm
King Systems Corp., Noblesville, IN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
10 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0418-2013 | KING LAD SILI CONE SINGLE USE, King Systems devices/case, ETO Sterile. Size 2 / Infant Silicone King LAD Patient Size: 10-20kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0419-2013 | KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD Patient Size: 30-50kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0420-2013 | KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 4 / Adult Silicone King LAD Patient Size: 50-70kg | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0421-2013 | KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 5 / Adult Silicone King LAD Patient Size: >70kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0422-2013 | KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie10 devices/case Size 3 / Child ClearSeal King LAD Patient Size: 30-50kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0423-2013 | KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 4 / Adult ClearSeal King LAD Patient Size: 50-70kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0424-2013 | KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD Patient Size: >70kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0425-2013 | KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 20 devices/case Size 2 / Infant ClearSeal Flexible King LAD Patient Size: 10-20kg Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0426-2013 | KING LAD Flexible ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case "Size 3 / Child ClearSeal Flexible King LAD Patient Size: 30-50kg " Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
| Z-0427-2013 | KING LAD Flexible ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case "Size 4 / Adult ClearSeal Flexible King LAD Patient Size: 50-70kg " Product Usage: Laryngeal airway | 263 cases of product. (2,630 devices) | Class II | Terminated | 28 Nov 2012 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2012
- 823
- Share of these events in Class II
- 88%
- Product records in this event
- 10Median of the firm: 1
- Events of the firm on the register
- 6
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 17 Feb 2015 | Double Swivel Connector with Suction Port Connector, Airway (Extension) King Systems part 7169 Product Usage: An airway | Device | Class II | 1 |
| 13 Dec 2013 | King Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airway Identification Part N | Device | Class II | 1 |
| 12 Dec 2013 | King LTS-D Oropharyngeal Airways Product Usage: The King LTS-D is intended for use in adult patients (in excess of 25 kg | Device | Class II | 1 |
| 3 Sep 2013 | King LT-D Oropharyngeal Airways The King LT-D is intended for use in adult patients (in excess of 25 kg) for controlled | Device | Class III | 1 |
| 23 Aug 2012 | Universal Flex2 Breathing Circuit Class I 510(k) Exempt An anesthesia breathing circuit is a device that is intended to | Device | Class I | 1 |
Official channels
- FDA enforcement report Z-0418-2013The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 63580. https://recallregister.com/fda/ii/63580/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
- What quantity of product is in FDA event 63580?
- The event has 10 product records. The first record gives this quantity:
263 cases of product. (2,630 devices)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2012 have the same class?
- 88% of the 823 device events that started in 2012 are in Class II.