Recall Register

Food, drugs, devices › King Systems Corp.

Device recall: KING LAD SILI CONE SINGLE USE, King Systems devices/case, ETO Sterile. Size 2 / Infant Silicone King LAD¿ Patient Size:

FDA enforcement event 63580 · Class II · initiated 2 Nov 2012 · status Terminated · terminated 28 May 2013

DeviceVoluntary: Firm initiated10 product recordsNoblesville, IN

FDA enforcement event 63580 is a device recall by King Systems Corp. of Noblesville, IN, initiated 2 Nov 2012 and classified Class II, with 10 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 823 device enforcement events initiated in 2012 carry the same class.

Status: Terminated, terminated 28 May 2013. Initiation: voluntary: firm initiated. King Systems Corp. has 6 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

King Systems Corp., Noblesville, IN.

Reason for recall

"The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO"

Distribution

"US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0418-2013KING LAD SILI CONE SINGLE USE, King Systems devices/case, ETO Sterile. Size 2 / Infant Silicone King LAD¿ Patient Size: 10-20kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0419-2013KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD¿ Patient Size: 30-50kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0420-2013KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 4 / Adult Silicone King LAD¿ Patient Size: 50-70kg263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0421-2013KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 5 / Adult Silicone King LAD¿ Patient Size: >70kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0422-2013KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie10 devices/case Size 3 / Child ClearSeal King LAD¿ Patient Size: 30-50kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0423-2013KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 4 / Adult ClearSeal King LAD¿ Patient Size: 50-70kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0424-2013KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD¿ Patient Size: >70kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0425-2013KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 20 devices/case Size 2 / Infant ClearSeal Flexible King LAD¿ Patient Size: 10-20kg Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0426-2013KING LAD Flexible ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case "Size 3 / Child ClearSeal Flexible King LAD¿ Patient Size: 30-50kg " Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012
Z-0427-2013KING LAD Flexible ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case "Size 4 / Adult ClearSeal Flexible King LAD¿ Patient Size: 50-70kg " Product Usage: Laryngeal airway263 cases of product. (2,630 devices)Class IITerminated28 Nov 2012

Context

88%of 2012 device events are Class II823 events that year
10product records in this eventfirm median: 1
6events on this register for the firm

Other enforcement events for King Systems Corp.

InitiatedProduct (first record)TypeClassRecords
17 Feb 2015Double Swivel Connector with Suction Port Connector, Airway (Extension) King Systems part 7169 Product Usage: An airway DeviceClass II1
13 Dec 2013King Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airway Identification Part NDeviceClass II1
12 Dec 2013King LTS-D Oropharyngeal Airways Product Usage: The King LTS-D is intended for use in adult patients (in excess of 25 kgDeviceClass II1
3 Sep 2013King LT-D Oropharyngeal Airways The King LT-D is intended for use in adult patients (in excess of 25 kg) for controlled DeviceClass III1
23 Aug 2012Universal Flex2 Breathing Circuit Class I 510(k) Exempt An anesthesia breathing circuit is a device that is intended to DeviceClass I1

All 6 events for King Systems Corp.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 28 May 2013. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.