Recall Register

Food, drugs, devices › Terumo Cardiovascular Systems Corporation

Device recall: System 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorpore

FDA enforcement event 62373 · Class II · initiated 25 Jun 2012 · status Terminated · terminated 1 Mar 2013

DeviceVoluntary: Firm initiated2 product recordsAnn Arbor, MI

Terumo Cardiovascular Systems Corporation (Ann Arbor, MI) initiated this device recall on 25 Jun 2012; FDA lists 2 product records under event 62373, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 823 device enforcement events initiated in 2012 carry the same class.

Status: Terminated, terminated 1 Mar 2013. Initiation: voluntary: firm initiated. Terumo Cardiovascular Systems Corporation has 73 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Terumo Cardiovascular Systems Corporation, Ann Arbor, MI.

Reason for recall

"Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu"

Distribution

"Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala, Honduras,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1970-2012System 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this…992 unitsClass IITerminated18 Jul 2012
Z-1971-2012System 1 Base 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this…648 unitsClass IITerminated18 Jul 2012

Context

88%of 2012 device events are Class II823 events that year
2product records in this eventfirm median: 2
73events on this register for the firm

Other enforcement events for Terumo Cardiovascular Systems Corporation

InitiatedProduct (first record)TypeClassRecords
28 May 2025CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial KDeviceClass II1
3 Mar 2025Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754DeviceClass II1
20 Feb 2024Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 6DeviceClass II1
1 Mar 2023CAPIOX RX15 Hollow Fiber Oxygenator, Catalog Number 3CX*RX15RE30DeviceClass II3
8 Jun 2021Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188DeviceClass II1
16 Apr 2021The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directlDeviceClass II3
2 Mar 2021Consist of a distal connection mechanism (either standard Terumo Ball Quick Connect or Terumo Shaft Quick Connect), an aDeviceClass II1
11 Nov 2020(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, Sterile Catalog NuDeviceClass II1
25 Aug 2020Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for useDeviceClass II3
10 Jan 2020Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for DeviceClass II4

All 73 events for Terumo Cardiovascular Systems Corporation

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala, Honduras,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 1 Mar 2013. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.