Recall Register

Food, drugs, devices › Cook Medical Incorporated

Device recall: Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZT

FDA enforcement event 97615 · Class I · initiated 18 Sep 2025 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordBloomington, IN

FDA enforcement event 97615 is a device recall by Cook Medical Incorporated of Bloomington, IN, initiated 18 Sep 2025 and classified Class I, with 1 product record.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 13% of the 945 device enforcement events initiated in 2025 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Cook Medical Incorporated has 8 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Cook Medical Incorporated, Bloomington, IN.

Reason for recall

"Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization."

Distribution

"Domestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC NJ NY OH OR PA SC TN TX UT VA WA WI WV International distribution to Switzerland, Germany, Italy, Poland, and Sweden"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0043-2026Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W ZTA2-P-42-173 ZTA2-P-42-173-W ZTA2-P-42-225 ZTA2-P-42-225-W ZTA2-P-44-125 ZTA2-P-44-125-W ZTA2-P-44-179 ZTA2-P-44-179-W ZTA2-P-44-233…695Class IOngoing22 Oct 2025

Context

13%of 2025 device events are Class I945 events that year
1product record in this eventfirm median: 2
8events on this register for the firm

Other enforcement events for Cook Medical Incorporated

InitiatedProduct (first record)TypeClassRecords
1 Apr 2024EchoTip¿ Ultra HD Endoscopic Ultrasound Access Needle, ECHO-HD-19-A, G52012DeviceClass II2
14 Feb 2022Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach Catalog Number/GPN: IGTCFS-65-1-FEM-TULIP G52917; IGTCFS-65DeviceClass III2
25 Feb 2019Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach, Catalog Number IGTCFS-65-1-FEM-TU LIP Product Usage: The GDeviceClass III6
25 Sep 2018Zilver 518 Biliary Stent 4mm x 20mm, Catalog Number: ZIB5-125-4.0-20 Product Usage: The Zilver 518 and 635 Biliary StentDeviceClass II39
22 Mar 2017Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascularDeviceClass I1
22 Dec 2015Zilver¿ 518RX Vascular Stent with Rapid Exchange Delivery SystemDeviceClass II1
26 Sep 2014Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that conDeviceClass II1

All 8 events for Cook Medical Incorporated

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Domestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC NJ NY OH OR PA SC TN TX UT VA WA WI WV International distribution to Switzerland, Germany, Italy, Poland, and Sweden"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.