Recall Register

Food, drugs, devices › Thoratec LLC

Device recall: HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with Hear

FDA enforcement event 96471 · Class I · initiated 13 Mar 2025 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordPleasanton, CA

Thoratec LLC (Pleasanton, CA) initiated this device recall on 13 Mar 2025; FDA lists 1 product record under event 96471, classified Class I.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 13% of the 945 device enforcement events initiated in 2025 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Thoratec LLC has 6 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Thoratec LLC, Pleasanton, CA.

Reason for recall

"Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power."

Distribution

"Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1570-2025HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.1661Class IOngoing23 Apr 2025

Context

13%of 2025 device events are Class I945 events that year
1product record in this eventfirm median: 2
6events on this register for the firm

Other enforcement events for Thoratec LLC

InitiatedProduct (first record)TypeClassRecords
13 May 2026Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US SoftwDeviceClass II3
10 Oct 2025CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-cDeviceClass II1
9 Oct 2025Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HearDeviceClass II4
23 Jun 2025Brand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Software Version: N/A DeviceClass I2
25 Jun 2024HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant KDeviceClass II1

All 6 events for Thoratec LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.