Device recall: Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environme
FDA enforcement event 96085 is a device recall by Fresenius Kabi USA, LLC of North Andover, MA, initiated 10 Jan 2025 and classified Class I, with 1 product record.
FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 13% of the 945 device enforcement events initiated in 2025 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Fresenius Kabi USA, LLC has 67 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Fresenius Kabi USA, LLC, North Andover, MA.
Reason for recall
"Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the…"
Distribution
"US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1019-2025 | Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration. | 23 systems | Class I | Ongoing | 12 Feb 2025 |
Context
Other enforcement events for Fresenius Kabi USA, LLC
All 67 events for Fresenius Kabi USA, LLC
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the…"
What does Class I mean?
FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
Where was the product distributed?
FDA's distribution pattern field reads: "US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI."
Is the recall still ongoing?