Recall Register

Food, drugs, devices › Thoratec Corp.

Device recall: HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2.

FDA enforcement event 93963 · Class I · initiated 19 Feb 2024 · status Ongoing

DeviceVoluntary: Firm initiated2 product recordsPleasanton, CA

FDA enforcement event 93963 is a device recall by Thoratec Corp. of Pleasanton, CA, initiated 19 Feb 2024 and classified Class I, with 2 product records.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 11% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Thoratec Corp. has 9 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Thoratec Corp., Pleasanton, CA.

Reason for recall

"Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System."

Distribution

"Worldwide distribution: U.S.(nationwide): AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1300-2024HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2. HeartMate II LVAS Implant Kit, US REF 106015 3. HeartMate II LVAS Implant Kit, OUS REF 106016 Update (February 14, 2024) HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II LVAS…2,240 kitsClass IOngoing27 Mar 2024
Z-1301-2024HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption (IDE) 2. HeartMate 3 Sealed Outflow Graft with Bend Relief, US 3. HeartMate 3 Sealed Outflow Graft with Bend Relief, OUS 4. HeartMate 3 LVAS Implant Kit, IDE 5. HeartMate 3 LVAS…22,663 kitsClass IOngoing27 Mar 2024

Context

11%of 2024 device events are Class I970 events that year
2product records in this eventfirm median: 1
9events on this register for the firm

Other enforcement events for Thoratec Corp.

InitiatedProduct (first record)TypeClassRecords
8 May 2024Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CANDeviceClass I1
1 Mar 2024Thoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnlyDeviceClass I2
3 Jan 20241. HeartMate Touch Communication System, REF HMT1100 2. HeartMate Touch Communication System, Rental REF HMT1100-R 3. HeDeviceClass I1
7 Aug 2023Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT; Heartmate XVE & HeartMate II Apical Coring Knife, REF: 1050DeviceClass II1
22 Oct 2021Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)DeviceClass II1
2 Dec 2019HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when coDeviceClass II1
30 Mar 2019Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist SystemDeviceClass II2
1 Mar 2019Thoratec HeartMate 3 Coring Tool, Sterile EO, Rx Only, REF 10005872, an optional accessory of Thoratec HeartMate 3 Left DeviceClass II1

All 9 events for Thoratec Corp.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution: U.S.(nationwide): AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.