Recall Register

Food, drugs, devices › Baxter Corporation Englewood

Device recall: ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Wo

FDA enforcement event 67884 · Class I · initiated 2 Apr 2014 · status Terminated · terminated 20 Jan 2016

DeviceVoluntary: Firm initiated1 product recordEnglewood, CO

Baxter Corporation Englewood (Englewood, CO) initiated this device recall on 2 Apr 2014; FDA lists 1 product record under event 67884, classified Class I.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 5% of the 1,176 device enforcement events initiated in 2014 carry the same class.

Status: Terminated, terminated 20 Jan 2016. Initiation: voluntary: firm initiated. Baxter Corporation Englewood has 9 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Baxter Corporation Englewood, Englewood, CO.

Reason for recall

"1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v"

Distribution

"Worldwide Distribution - USA nationwide, Canada, Colombia, Dominican Republic, Panama, Austria, Bosnia Herzegovina, Bulgaria, Bahrain, Estonia, France, Germany, Georgia, Hungary, Ireland, Israel,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1566-2014ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition), 8300-0156 ABACUS 3.0 CE, 8300-0157 ABACUS 3.0 SE, 8300-0158 ABACUS 3.0 ME, 8300-0166 ABACUS 3.1 DE, 8300-0167 ABACUS 3.1 CE,…2034Class ITerminated21 May 2014

Context

5%of 2014 device events are Class I1,176 events that year
1product record in this eventfirm median: 1
9events on this register for the firm

Other enforcement events for Baxter Corporation Englewood

InitiatedProduct (first record)TypeClassRecords
24 Aug 20160.2 Micron Filter, 50 mm Product Usage: The 0.2 Micron Filter is a bacteria and particulate filter for aqueous solutionsDeviceClass I1
21 Dec 2015Baxter, RAPIDFILL Syringe Strip, 10 ml. For use with RAPIDFILL Automated Syringe Filler (ASF)DeviceClass II1
8 Apr 2015ABACUS SE (Single-Workstation Edition) and ABACUS ME (Multi-Workstation Edition) Product Usage: The ABACUS Software is aDeviceClass II1
29 Dec 2014Baxter, Self-Righting Syringe Tip Caps, Self-Righting DISCPAC Syringe Tip Caps. Product codes H93866100, H93869100, H938DeviceClass II1
4 Dec 2014Self-Righting Luer Lock Tip Cap, Green and Self-Righting Luer Slip Tip Cap, Yellow, Baxter Healthcare Corporation, 14445DeviceClass II1
28 May 2014Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.DeviceClass II1
16 Apr 2014RAPIDFILL Syringe Strip, For use with RAPIDFILL Automated Syringe Filler (ASF): H93890200 RAPIDFILL Syringe Strip, WhiteDeviceClass II1
3 Dec 2013ABACUS TPN Calculation Software, Baxa: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-WorkstationDeviceClass II1

All 9 events for Baxter Corporation Englewood

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v"

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - USA nationwide, Canada, Colombia, Dominican Republic, Panama, Austria, Bosnia Herzegovina, Bulgaria, Bahrain, Estonia, France, Germany, Georgia, Hungary, Ireland, Israel,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 20 Jan 2016. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.